Reaction calorimeter
Laboratory
SPECIFICATIONS
- Manufacturer : METTLER TOLEDO
- Model : OptIMax™ HFCal
- Thermostat :
- Heating: Electrical
- Cooling: Peltier
- Temperature range : -40°C to 180°C
- Control Modes : Isothermal and isoperibolic, constant or ramp, reflux, distillation and crystallization
- Stirring : 30 rpm to 1200 rpm
- Reactors :
- Type : Single piece or two piece
- Volume : 1000, 500 and 250 ml
- Pressure : 50 mbar to 1 bar
- Material : Duran® glass
- Stirrer : Pitch-blade, Anchor, Half-moon
- Calorimetry :
- Type : Heat flow
- Precision Heat Transfer : Typically ± 3 %
- Accuracy and Precision Specific Heat : Typically ± 10 %
- Accuracy Heat Flow : Typically ± 3 to 5 % (under isothermal conditions); Typically ± 5 to 10 % (under non-isothermal conditions)
PARTICULARITY
- Accessories: Modules for dose addition, pH probe, turbidity probe
USE
- Determining how chemical reactions can be transferred safety from the laboratory to the production site.
- OptIMax™ HFCal reproducibly delivers information to characterize process parameters, for scale-up and to ensure it is safe at scale:
- Heat transfer coefficient
- Specific heat of the reaction mass
- Heat flow for isothermal and non-isothermal conditions
- Enthalpy
- Thermal conversion
- Adiabatic temperature rise